Calibration factors for dosemeter characterization

Description

Calibration factors or dosemeter sensitivities can be used for linking the measured physical effect with the biological effect. Calibration factors have been determined for a solid-state track detector system (cellulose triacetate foil and LiF) in different neutron fields (252Cf fission neutrons, WWR-S reactor neutrons with and without Cd and Fe shielding, (d,d) generator neutrons, U-120 cyclotron neutrons with and without Fe and paraffin shielding, 9.2 MeV neutrons, and (d,t) generator neutrons). The influence of physical factors on the calibration factor and on the measuring range and measuring accuracy is demonstrated and discussed

Availability note (English)

MF available from INIS under the Report Number.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Additional titles

Original title (German)
Der Kalibrierungsfaktor zur Charakterisierung von Dosimetern

Publishing Information

Imprint Pagination
8 p.
Report number
TU-Inf--05-06-81