Published 1981
| Version v1
Report
Restricted
Calibration factors for dosemeter characterization
Description
Calibration factors or dosemeter sensitivities can be used for linking the measured physical effect with the biological effect. Calibration factors have been determined for a solid-state track detector system (cellulose triacetate foil and LiF) in different neutron fields (252Cf fission neutrons, WWR-S reactor neutrons with and without Cd and Fe shielding, (d,d) generator neutrons, U-120 cyclotron neutrons with and without Fe and paraffin shielding, 9.2 MeV neutrons, and (d,t) generator neutrons). The influence of physical factors on the calibration factor and on the measuring range and measuring accuracy is demonstrated and discussed
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Additional titles
- Original title (German)
- Der Kalibrierungsfaktor zur Charakterisierung von Dosimetern
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- TU-Inf--05-06-81
INIS
- Country of Publication
- German Democratic Republic
- Country of Input or Organization
- German Democratic Republic
- INIS RN
- 14788466
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CALIBRATION; DIELECTRIC TRACK DETECTORS; DOSEMETERS; ERRORS; FAST NEUTRONS; INTERMEDIATE NEUTRONS; LITHIUM FLUORIDES; NEUTRON DOSIMETRY; NEUTRON FLUENCE; NEUTRON SPECTRA; OPTICAL FILTERS; RADIATION DOSES; RESPONSE FUNCTIONS; SENSITIVITY; THERMAL NEUTRONS
- Descriptors DEC
- ALKALI METAL COMPOUNDS; BARYONS; DOSIMETRY; ELEMENTARY PARTICLES; FERMIONS; FILTERS; FLUORIDES; FLUORINE COMPOUNDS; FUNCTIONS; HADRONS; HALIDES; HALOGEN COMPOUNDS; LITHIUM COMPOUNDS; LITHIUM HALIDES; MEASURING INSTRUMENTS; NEUTRONS; NUCLEONS; RADIATION DETECTORS; SPECTRA