Published August 1989 | Version v1
Report Open

Calculation of neutron detection efficiency for the thick lithium glass using Monte Carlo method

  • 1. Beijing Univ. (China). Inst. of Heavy Ion Physics
  • 2. Beijing Nuclear Instrument Factory, BJ (China)

Description

The neutron detector efficiencies of a NE912 (45mm in diameter, 9.55 mm in thickness) and 2 pieces of ST601 (40mm in diameter, 3 and 10 mm in thickness respectively) lithium glasses have been calculated with a Monte Carlo computer code. The energy range in the calculation is 10 keV to 2.0 MeV. The effect of time delayed caused by neutron multiple scattering in the detectors (prompt neutron detection efficiency) has been considered

Availability note (English)

MF available from INIS under the Report Number.

Files

21090769.pdf

Files (313.0 kB)

Name Size Download all
md5:5d6e1aa0e9c3c217a0b523bc1433020e
313.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
9 p.
Report number
CNIC--00316

Optional Information

Secondary number(s)
PU--0003.