An in-phantom dosimetry system using pin silicon photodiode radiation sensors for measuring organ doses in x-ray CT and other diagnostic radiology
- 1. School of Health Sciences, Nagoya University, Daikominami, Higashi-ku, Nagoya 461-8673 (Japan)
Description
A dosimetry system using commercially available pin silicon photodiodes as the sensor is evaluated for in-phantom dose measurements in x-ray CT and other diagnostic radiology. System sensitivity measured as a function of the effective energy of x rays was between 0.37 and 0.49 V/mGy at an effective energy range between 23.5 and 72 keV. The minimum detectable organ dose with 25% uncertainty was estimated to be 0.02 mGy. The excellent output linearity was found over a dose range from 0.03 to more than 10 mGy with flat dose rate response of system sensitivity up to 35 mGy s-1, though the sensitivity indicated some energy dependence across the diagnostic energy range with a maximum of about 10%/10 keV. Since angular dependence of the sensitivity of the photodiode sensor was found to be small enough it would induce negligible dose error. Dose profile measurement along the axis of a thoracic phantom undergoing CT chest examination indicated the reliability of dose values over a range of two orders of magnitude from less than 0.2 to 12 mGy. The present dosimetry system having advantages of high sensitivity with immediate readout of dose values, low cost, and easy construction would widely be used as an alternative to TLD dosimeters for organ and skin dose measurements in CT and other diagnostic radiology
Additional details
Identifiers
- DOI
- 10.1118/1.1489042;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 29
- Journal Issue
- 7
- Journal Page Range
- p. 1504-1510
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 35004285
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CHEST; COMPUTERIZED TOMOGRAPHY; PHANTOMS; RADIATION DETECTORS; RADIATION DOSES; SILICON DIODES; SKIN; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; DOSIMETRY; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MOCKUP; ORGANS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- (c) 2002 American Association of Physicists in Medicine.