Published 1992 | Version v1
Report Open

Fiber-optic pH sensor for in-situ applications

  • 1. CEA Centre d'Etudes de la Vallee du Rhone, 30 - Marcoule (France). Dept. des Procedes de Retraitement
  • 2. CEA Centre d'Etudes de Fontenay-aux-Roses, 92 (France)

Description

An evaluation of the feasibility of a fiber-optic pH sensor was carried out. The pH sensor is composed of an optode connected to the transmitter-receiver apparatus (Optolec H) through a 100 m-long fiber. Three diodes of carefully chosen wavelengths are located in the Optolec H and emit through a bundle of nine fibers to the end of the optode where the oncoming light is reflected on an immobilized pH-indicator-based spherical bead of resin and sent back for analysis to the Optolec H through a central fiber. The influence of ionic strength as well as precision, response time, and lifetime have been determined. This device is to be used for routine in-situ measurements in an underground laboratory devoted to studies in connection with nuclear waste repositories where harsh conditions are encountered for direct collection of waters and traditional post-determination with a pH-meter

Availability note (English)

MF available from INIS under the Report Number.

Files

24007705.pdf

Files (258.8 kB)

Name Size Download all
md5:e4f2b7639bbda2891722e025c6bd4fbf
258.8 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
21 p.
Report number
CEA-CONF--11072

Conference

Title
1. European Conference on Optical Chemical Sensors and Biosensors.
Dates
12-15 Apr 1992.
Place
Graz (Austria).