Published May 1994 | Version v1
Journal article

Evaluation of Drung-type magnetometers for multi-channel systems

  • 1. Superconducting Sensor Lab., Chiba (Japan)

Description

The properties of a prototype Drung-type SQUID with single-digit or smaller dynamic resistance, decreased using asymmetric bias current injection, are examined. Magnetic field resolution of 2.2 fT Hz-1/2 at 1 kHz and high sensitivity are obtained. A highly reliable ceramic package is developed for multi-channel applications. A prototype 256-channel system for biomagnetic application using mass-produced SQUIDS is built, and the successful operation of 206 channels is verified. (author)

Availability note (English)

Available online at the Web site for the journal Superconductor Science and Technology (ISSN 1361-6668) http://www.iop/org/

Additional details

Identifiers

Publishing Information

Journal Title
Superconductor Science and Technology
Journal Volume
7
Journal Issue
5
Journal Page Range
p. 249-252
ISSN
0953-2048

Conference

Title
International superconductive electronics conference
Dates
Aug 1993
Place
Boulder, CO (United States)