A rapid, accurate and highly sensitive method for the determination of radon in water with a liquid scintillation counter is described. The method is based on the high solubility of radon in toluene (toluene scintillator solution). An overall recovery of 90.5% (including the extraction and detection processes) was achieved; the limits of detection and determination were 0.054 and 0.12 Bq/l, respectively. The total error of determination did not exceed 10% for radioactivities not lower than 1 Bq/l. (Z.S.) 1 tab., 3 figs., 6 refs

Part of:
Radionuclides and ionizing radiation in water management

Additional details

Additional titles

Original title (Czech)
Stanoveni radonu ve vode metodou LSC

Publishing Information

Publisher
Ceska Vedeckotechnicka Vodohospodarska Spolecnost.
Imprint Place
Ostrava (Czech Republic)
ISBN
80-02-00986-X
Imprint Title
Radionuclides and ionizing radiation in water management
Imprint Pagination
286 p.
Journal Page Range
p. 169-175.
Report number
INIS-mf--14411

Conference

Title
14. conference on radionuclides and ionizing radiation in water management.
Original Conference Title
XIV. konference 'Radionuklidy a ionizujici zareni ve vodnim hospodarstvi'
Dates
26-27 Oct 1993.
Place
Lazne Bohdanec (Czech Republic).

Optional Information

Notes
Imprint:Radionuklidy a ionizujici zareni ve vodnim hospodarstvi.