Quantitative radiographic imaging using a photostimulable phosphor system
- 1. Duke Univ. Medical Center, Durham, NC (USA)
Description
We have evaluated a photostimulable phosphor x-ray imaging system [Philips Computed Radiography (PCR) system] for use in quantification of x-ray exposure in diagnostic radiography. An exponential function was fitted to data yielding quantitative x-ray exposure values as a function of digital pixel values. We investigated several factors that affect the accuracy of exposure measurement using the PCR including repeatability, background noise as a function of time delay between plate erasure and use, sensitivity variation between different plates, nonuniformity of sensitivity within a plate, decay of the latent image between time of exposure and readout (observed as a change in sensitivity), and the accuracy with which the (exponential) calibration function yields exposure values as a function of digital pixel values. The calibration was performed over the exposure range from 5.1 X 10(-9) to 2.0 X 10(-5) C/kg (0.02-75 mR). The accuracy of exposure measurements made with a single imaging plate is between 1.6% and 4.2%. If measurements from several plates are involved, the uncertainty in the final measurement will be between 5% and 5.9%
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 17
- Journal Issue
- 3
- Series
- Med. Phys.
- Journal Page Range
- 454-459
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 22008961
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CALIBRATION; EVALUATION; RADIATION DOSES; SENSITIVITY ANALYSIS; X RADIATION
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; RADIATIONS