Quantitative radiographic imaging using a photostimulable phosphor system

  • 1. Duke Univ. Medical Center, Durham, NC (USA)

Description

We have evaluated a photostimulable phosphor x-ray imaging system [Philips Computed Radiography (PCR) system] for use in quantification of x-ray exposure in diagnostic radiography. An exponential function was fitted to data yielding quantitative x-ray exposure values as a function of digital pixel values. We investigated several factors that affect the accuracy of exposure measurement using the PCR including repeatability, background noise as a function of time delay between plate erasure and use, sensitivity variation between different plates, nonuniformity of sensitivity within a plate, decay of the latent image between time of exposure and readout (observed as a change in sensitivity), and the accuracy with which the (exponential) calibration function yields exposure values as a function of digital pixel values. The calibration was performed over the exposure range from 5.1 X 10(-9) to 2.0 X 10(-5) C/kg (0.02-75 mR). The accuracy of exposure measurements made with a single imaging plate is between 1.6% and 4.2%. If measurements from several plates are involved, the uncertainty in the final measurement will be between 5% and 5.9%

Additional details

Publishing Information

Journal Title
Medical Physics
Journal Volume
17
Journal Issue
3
Series
Med. Phys.
Journal Page Range
454-459
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
22008961
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; CALIBRATION; EVALUATION; RADIATION DOSES; SENSITIVITY ANALYSIS; X RADIATION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; RADIATIONS