Published January 1990 | Version v1
Journal article

Film dosimetry of clinical electron beams

  • 1. Dr. Daniel den Hoed Cancer Centre, Rotterdam (Netherland)

Description

Film dosimetry is used widely to obtain relative dose distributions of clinical electron beams in phantoms. Nevertheless, measurement results obtained with film dosimetry may lack precision and reliability. In this paper well defined and reproducible methods in film dosimetry are discussed. By application of these methods, film dosimetry appears to be adequate in measuring relative dose distributions of clinically applied electron beams, with an accuracy of 1% to 2% of the dose maximum, in water and plastics as well as in heterogeneously composed material

Additional details

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
18
Journal Issue
1
Series
Int. J. Radiat. Oncol., Biol. Phys.
Journal Page Range
69-76
ISSN
0360-3016
CODEN
IOBPD

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
21053356
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ACCURACY; ELECTRON BEAMS; FILM DOSIMETRY; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
BEAMS; DOSIMETRY; LEPTON BEAMS; MEDICINE; MOCKUP; PARTICLE BEAMS; STRUCTURAL MODELS; THERAPY