Published January 1990
| Version v1
Journal article
Film dosimetry of clinical electron beams
Creators
- 1. Dr. Daniel den Hoed Cancer Centre, Rotterdam (Netherland)
Description
Film dosimetry is used widely to obtain relative dose distributions of clinical electron beams in phantoms. Nevertheless, measurement results obtained with film dosimetry may lack precision and reliability. In this paper well defined and reproducible methods in film dosimetry are discussed. By application of these methods, film dosimetry appears to be adequate in measuring relative dose distributions of clinically applied electron beams, with an accuracy of 1% to 2% of the dose maximum, in water and plastics as well as in heterogeneously composed material
Additional details
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 18
- Journal Issue
- 1
- Series
- Int. J. Radiat. Oncol., Biol. Phys.
- Journal Page Range
- 69-76
- ISSN
- 0360-3016
- CODEN
- IOBPD
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21053356
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCURACY; ELECTRON BEAMS; FILM DOSIMETRY; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- BEAMS; DOSIMETRY; LEPTON BEAMS; MEDICINE; MOCKUP; PARTICLE BEAMS; STRUCTURAL MODELS; THERAPY