Published April 1, 2018 | Version v1
Journal article

Impact of the positive ion current on large size neutrino detectors and delayed photon emission

  • 1. Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas (CIEMAT), Av. Complutense 40, 28040 Madrid (Spain)
  • 2. Institut de Física d´Altes Energies (IFAE), Campus UAB, Bellaterra (Barcelona) (Spain)

Description

Given their small mobility coefficient in liquid argon with respect to the electrons, the ions spend a considerably longer time in a sensitive detector volume instrumented with an electric field. We studied the effects of the positive ion current in a liquid argon time projection chamber, in the context of massive argon experiments for neutrino physics. The constant recombination between free ions and electrons produces a quenching of the charge signal and a constant emission of photons, uncorrelated in time and space to the physical interactions. The predictions evidence some potential concerns for multi-ton argon detectors, particularly when operated on surface.

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/13/04/C04015

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
13
Journal Issue
04
Journal Page Range
p. C04015
ISSN
1748-0221