A study on characteristic of glass dosimeter according to grade change of tube current

  • 1. Dept. of Radiological Science, Shingu University, Sungnam (Korea, Republic of)
  • 2. Dept. of Radiology, Samsung Medical Center, Seoul (Korea, Republic of)

Description

This study was evaluated the linearity and reproducibility according to dose, and reproducibility according to delay time by changing tube current amount (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively, which are low energy radiations) using Glass Dosimeter (GD) and piranha semiconductor dosimeter which are used for measuring exposure dose. Measurements of radiation dose were performed using external detector of piranha 657 which is multi-function QA device (RTI Electronic, Sweden). Conditions of measurement were 80 kVp, SSD 100 and exposure region is 10 cm x 10 cm. Glass dosimeter was exposed to radiation. Twenty-four glass dosimeters were divided into six groups (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively), then measured. This study was resulted by measuring the linearity and reproducibility according to change of tube current in low energy field. In dose characteristic of GD, this study could be useful as previous study with regard to dose characteristic according to change of tube voltage in low energy field

Additional details

Publishing Information

Journal Title
Journal of the Korean Society of Radiological Technology
Journal Volume
37
Journal Issue
2
Series
16 refs, 2 figs, 5 tabs
Journal Page Range
p. 135-141
ISSN
1226-2854

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
46110771
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSEMETERS; GLASS; RADIATION DOSES; RADIATIONS; TUBES
Descriptors DEC
DOSES; MEASURING INSTRUMENTS