A study on characteristic of glass dosimeter according to grade change of tube current
Creators
- 1. Dept. of Radiological Science, Shingu University, Sungnam (Korea, Republic of)
- 2. Dept. of Radiology, Samsung Medical Center, Seoul (Korea, Republic of)
Description
This study was evaluated the linearity and reproducibility according to dose, and reproducibility according to delay time by changing tube current amount (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively, which are low energy radiations) using Glass Dosimeter (GD) and piranha semiconductor dosimeter which are used for measuring exposure dose. Measurements of radiation dose were performed using external detector of piranha 657 which is multi-function QA device (RTI Electronic, Sweden). Conditions of measurement were 80 kVp, SSD 100 and exposure region is 10 cm x 10 cm. Glass dosimeter was exposed to radiation. Twenty-four glass dosimeters were divided into six groups (5 mAs, 10 mAs, 16 mAs, 20 mAs, 25 mAs, 32 mAs respectively), then measured. This study was resulted by measuring the linearity and reproducibility according to change of tube current in low energy field. In dose characteristic of GD, this study could be useful as previous study with regard to dose characteristic according to change of tube voltage in low energy field
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society of Radiological Technology
- Journal Volume
- 37
- Journal Issue
- 2
- Series
- 16 refs, 2 figs, 5 tabs
- Journal Page Range
- p. 135-141
- ISSN
- 1226-2854
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 46110771
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DOSEMETERS; GLASS; RADIATION DOSES; RADIATIONS; TUBES
- Descriptors DEC
- DOSES; MEASURING INSTRUMENTS