Published February 1990 | Version v1
Journal article

CaSO4:Dy environmental TLD performance testing

  • 1. Duke Power Co., Charlotte, NC (United States)

Description

This paper reports that this study involved performance testing of the Teledyne Isotopes calcium sulfate dysprosium (CaSO4:Dy) environmental thermoluminescent dosimeter (TLD), and Model PB-2 badge case against specific portions of the American National Standards Institute N545-1975 (ANSI N545) standard for environmental TLD applications. Dosimeters were tested in seven categories of ANSI N545, ranging from reproducibility to self-irradiation. Dosimeters in the tests demonstrate the stable, accurate, and predictable behavior of CaSO4:Dy for environmental radiation monitoring, as shown in the discussion for each test category. Specific observations are made on environmental shielding effects demonstrated in the moisture and self-irradiation tests

Additional details

Publishing Information

Journal Title
Radiation Protection Management
Journal Volume
7
Journal Issue
1
Series
Radiat. Prot. Manage.
Journal Page Range
68-78
ISSN
0740-0640
CODEN