Published January 1997 | Version v1
Journal article

Alanine EPR dosimeter response in proton therapy beams

  • 1. Massachusetts General Hospital, Boston, MA (United States)
  • 2. National Inst. of Standards and Technology (NML), Gaithersburg, MD (United States). Ionizing Radiation Div.

Description

We report a series of measurements directed to assess the suitability of alanine as a mailable dosimeter for dosimetry quality assurance of proton radiation therapy beams. These measurements include dose-response of alanine at 140 MeV, and comparison of response vs energy with a parallel plate ionization chamber. All irradiations were made at the Harvard Cyclotron Laboratory, and the dosimeters were read at NIST. The results encourage us that alanine could be expected to serve as a mailable dosimeter with systematic error due to differential energy response no greater than 3% when doses of 25 Gy are used. (Author)

Additional details

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
47
Journal Issue
11-12
Journal Page Range
p. 1197-1199.
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
4. international symposium on ESR dosimetry and applications.
Dates
15-19 May 1995.
Place
Munich (Germany).