Published February 7, 2009 | Version v1
Journal article

Small field measurements with a novel silicon position sensitive diode array

  • 1. Rutherford Appleton Laboratory, Science and Technology Facilities Council, Chilton OX110QX (United Kingdom)
  • 2. Department of Medical Physics, University Hospital Birmingham NHS Trust, Birmingham B15 2TH (United Kingdom)
  • 3. School of Medicine, Swansea University, Swansea SA2 8PP (United Kingdom)
  • 4. Physics and Astronomy Dept., University of Birmingham, Birmingham B15 2TT (United Kingdom)

Description

DOSI, a novel dosimeter based on position sensitive detectors for particle physics experiments, was used for relative clinical dosimetry measurements in small radiotherapy fields. The device is capable of dynamic measurements in real time and provides sub-millimetre spatial resolution. The basic beam data for a stereotactic radiotherapy collimator system (BrainLAB) using 6 MV photons were measured and compared with the corresponding data acquired with a small diamond detector and a PinPoint ionization chamber. All measurements showed an excellent agreement between DOSI and the diamond detector. There was an increasing discrepancy between the relative output factors (ROF) measured with DOSI and those measured with the ionization chamber with decreasing field size, specifically for collimators with a diameter smaller than 15 mm. The percentage depth doses (PDD) were in agreement to better than 1% for all depths. The agreement on off-axis ratios (OAR) was better than 3% for all collimators, whereas the agreement on relative output factors (ROF) was at the 1% level. DOSI's fast read-out electronics made it possible for all measurements to be recorded within 45 min including time to change collimators. This should reduce the overall time for commissioning and QA measurements, an important factor especially for busy radiotherapy departments.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/54/3/002

Additional details

Identifiers

DOI
10.1088/0031-9155/54/3/002;
PII
S0031-9155(09)83427-1;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
54
Journal Issue
3
Journal Page Range
p. 485-495
ISSN
0031-9155
CODEN
PHMBA7