DOSI, a novel dosimeter based on position sensitive detectors for particle physics experiments, was used for relative clinical dosimetry measurements in small radiotherapy fields. The device is capable of dynamic measurements in real time and provides sub-millimetre spatial resolution. The basic beam data for a stereotactic radiotherapy collimator system (BrainLAB) using 6 MV photons were measured and compared with the corresponding data acquired with a small diamond detector and a PinPoint ionization chamber. All measurements showed an excellent agreement between DOSI and the diamond detector. There was an increasing discrepancy between the relative output factors (ROF) measured with DOSI and those measured with the ionization chamber with decreasing field size, specifically for collimators with a diameter smaller than 15 mm. The percentage depth doses (PDD) were in agreement to better than 1% for all depths. The agreement on off-axis ratios (OAR) was better than 3% for all collimators, whereas the agreement on relative output factors (ROF) was at the 1% level. DOSI's fast read-out electronics made it possible for all measurements to be recorded within 45 min including time to change collimators. This should reduce the overall time for commissioning and QA measurements, an important factor especially for busy radiotherapy departments.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/3/002Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/3/002;
- PII
- S0031-9155(09)83427-1;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 3
- Journal Page Range
- p. 485-495
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41013411
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; DEPTH DOSE DISTRIBUTIONS; DIAMONDS; IONIZATION CHAMBERS; PHOTONS; POSITION SENSITIVE DETECTORS; RADIOTHERAPY; SILICON
- Descriptors DEC
- BOSONS; CARBON; ELEMENTARY PARTICLES; ELEMENTS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; MEDICINE; MINERALS; NONMETALS; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SEMIMETALS; SPATIAL DOSE DISTRIBUTIONS; THERAPY