Description
The analysis was performed of corrections for dead time and false coincidence which had been used in calculating activities in the 4π β-γ coincidence method. In the correction analysis the Monte Carlo method was used facilitating to clear up probable inaccuracies of the used corrections. It was shown experimentally that using the equipment described specimens cannot be measured with activities exceeding 104 Bq owing to the incorrect function of the single-channel analyzer in the gamma channel at high activities. (author)
Availability note (English)
Available from Ustav pro vyzkum, vyrobu a vyuziti radioisotopu, Prague, Czechoslovakia.Additional details
Additional titles
- Original title (Czech)
- Overeni funkce 4π β-γ koincidencniho zarizeni pro velmi vysoke aktivity
Publishing Information
- Imprint Pagination
- 60 p.
- Report number
- CS-INIS--194
INIS
- Country of Publication
- Serbia and Montenegro
- Country of Input or Organization
- Serbia and Montenegro
- INIS RN
- 9405248
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- ANALYTICAL SOLUTION; BETA SPECTROSCOPY; COINCIDENCE SPECTROMETRY; CORRECTIONS; DEAD TIME; ERRORS; FOUR-PI COUNTING; GAMMA SPECTROSCOPY; MONTE CARLO METHOD; PERFORMANCE TESTING; PULSE ANALYZERS; TIME RESOLUTION
- Descriptors DEC
- COINCIDENCE METHODS; COUNTING TECHNIQUES; ELECTRONIC EQUIPMENT; RESOLUTION; SPECTROSCOPY; TESTING; TIMING PROPERTIES