By independent research LⅢ--absorption edge densimeter, an analytical method for uranium sample with the concentration from 20 g/L to 200 g/L was developed. The fitting area for uranium measurement was determined through experiment. The left fitting area was 1659-1856 channel, and the right one was 2063-2280 channel. The uranium LⅢ--absorption edge was at 1995 channel. The results show that the influence of HNO3 concentration lower than 9 mol/L, Al and Fe concentration lower than 10 g/L was negligibly small. The uranium measurement precision is better than 0.5%, and the instrument stability is good. Some samples in bench test of uranium recovery were determined. The results are satisfactory. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- Atomic Energy Science and Technology
- Journal Volume
- 48
- Journal Issue
- 8
- Journal Page Range
- p. 1351-1355
- ISSN
- 1000-6931
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 48088578
- Subject category
- S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Descriptors DEI
- ABSORPTION; ABSORPTION SPECTRA; ACCURACY; BENCH-SCALE EXPERIMENTS; CONCENTRATION RATIO; DENSIMETERS; ENERGY SPECTRA; MATERIALS RECOVERY; NITRIC ACID; STABILITY; URANIUM; X-RAY SPECTRA
- Descriptors DEC
- ACTINIDES; DIMENSIONLESS NUMBERS; ELEMENTS; HYDROGEN COMPOUNDS; INORGANIC ACIDS; INORGANIC COMPOUNDS; MANAGEMENT; MEASURING INSTRUMENTS; METALS; NITROGEN COMPOUNDS; OXYGEN COMPOUNDS; PROCESSING; SORPTION; SPECTRA; WASTE MANAGEMENT; WASTE PROCESSING
Optional Information
- Notes
- 6 figs., 4 tabs., 13 refs.; http://dx.doi.org/10.7538/yzk.2014.48.08.1351