Published November 2018 | Version v1
Journal article

Performance evaluation of axxent electronic brachytherapy system model 110

  • 1. Radiological Safety Division, Atomic Energy Regulatory Board, Mumbai (India)

Description

Traditional brachytherapy refers to the placement of radioactive sources on or inside the tumor. Electronic brachytherapy (EB) is new form of radiotherapy which utilizes miniature X-ray source instead of radioactive source to deliver desired doses to tumor. The advantages of EB include low dose to surrounding organs, simple emergency handling procedures, lesser shielding requirements, no leakage radiation in off state and no radioactive waste. Purpose of this paper is to discuss performance evaluation of Axxent EB system model 110 with respect to conformance with manufacturer's specifications and existing QA protocol of brachytherapy, as applicable. This unit is intended to deliver intracavitary, surface radiation and Intra Operative Radiation Therapy (IORT) and not suitable for interstitial application due to larger source dimension

Additional details

Publishing Information

Journal Title
Journal of Medical Physics
Journal Volume
43
Journal Issue
suppl.1
Journal Page Range
p. 28-29
CODEN
JMPHFE

Conference

Title
39. annual conference of Association of Medical Physicists of India
Acronym
AMPICON 2018
Dates
2-4 Nov 2018
Place
Chennai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
50065526
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BRACHYTHERAPY; DOSE LIMITS; DOSIMETRY; NEOPLASMS; QUALITY ASSURANCE; TUMOR CELLS; X-RAY SOURCES
Descriptors DEC
ANIMAL CELLS; DISEASES; MANAGEMENT; MEDICINE; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIATION SOURCES; RADIOLOGY; RADIOTHERAPY; SAFETY STANDARDS; STANDARDS; THERAPY