Evaluation of polymer-coated CsI:Tl as an alpha/beta pulse shape discriminating flow-cell

  • 1. Nuclear Regulatory Commission, Washington, DC (United States)
  • 2. Clemson Univ., SC (United States)

Description

A pulse shape discriminating flow-cell radiation detection system constructed with polymer coated CsI:Tl was evaluated for simultaneous gross alpha/gross beta quantification. The CsI:TI scintillator was crushed and sieved to 63-90 μm particle size and microencapsulated with Parylene C to reduce its rate of dissolution. Averaged over the first hour of use, the pulse shape discrimination figure-of-merit was 1.4 and the detection efficiencies were 64.9 ± 5.7 %, 52.5 ± 4.5 % and 4.5 ± 0.2 % for 233U, 90Sr/90Y and 14C , respectively. The typical background count rate in the alpha and beta pulse shape window was 0.17 and 0.004 cps, respectively. The resultant minimum detectable activity for a 30 second count time was calculated to be 0.19 ± 0.01 Bq, 0.9 ± 0.1 Bq and 11.4 ± 0.6 Bq for 233U, 90Sr/90Y and 14C, respectively. Although the 3 μm thick microencapsulation reduced CsI:Tl dissolution, the detection efficiency declined by a factor of two after 4.8 hours while the pulse shape resolution degraded slightly

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 865-869.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

Optional Information

Contract/Grant/Project number
Contract AR21-95MC32110
Secondary number(s)
CONF-961123--.