Published 2007 | Version v1
Journal article

Simulation studies on a prototype ionisation chamber for measurement of personal dose equivalent, Hp(10)

  • 1. Instituto Tecnologico e Nuclear, Departamento de Protecao Radiologica e Seguranca Nuclear, Estrada Nacional 10, 2686-953 Sacavem (Portugal)

Description

A prototype ionisation chamber for direct measurement of the personal dose equivalent, Hp(10), similar to the one developed by the Physikalisch-Technische Bundesantalt (PTB), was designed and constructed by the Metrological Laboratory of Ionizing Radiation (LMRI) of Nuclear and Technological Inst. (ITN). Tests already performed have shown that the behaviour of this chamber is very similar to the PTB chamber, mainly the energy dependence for the X-ray radiation qualities of the ISO 4037-1 narrow series N-30, N-40, N-60, N-80, N-100 and N-120 and also for gamma radiation of 137Cs and 60Co. However, the results obtained also show a dependence on the energy and angles of incident radiation and a low magnitude of the electrical response of the ionisation chamber. In order to optimise the performance of the chamber, the LMRI initiated numerical simulation of this ionisation chamber by Monte Carlo method using the MCNPX code. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncm207

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
125
Journal Issue
1-4
Journal Page Range
p. 175-179
ISSN
0144-8420

Conference

Title
European Workshop on Individual Monitoring of Ionising Radiation
Acronym
IM2005
Dates
11-15 Apr 2005
Place
Vienna (Austria)

Optional Information

Notes
7 refs