Published 1987 | Version v1
Book

Liquid-argon ionization chambers as total absorption detectors

Description

A new detector for the measurement of energy by total absorption, based on the use of multiple-plate ion chambers, is described. The use of liquid argon as the working medium and optimized readout results in an electronic noise contribution to the resolution of less than 0.1 GeV, in a large detector. The use of thin plates, 0.1 radiation length, ensure that sampling fluctuations are small. The technique allows absolute calibration and very good gain stability. Tests on a detector large enough to absorb a high-energy electromagnetic shower are described, where the energy resolution is limited by the residual sampling fluctuations

Additional details

Publishing Information

Publisher
Addison-Wesley Pub. Co. Inc.
Imprint Place
Reading, MA (USA)
Imprint Title
Experimental techniques in high energy physics
Journal Page Range
p. 497-534.

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
20035101
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ARGON; CALIBRATION; FLUCTUATIONS; IONIZATION CHAMBERS; PLATES; SAMPLING
Descriptors DEC
ELEMENTS; MEASURING INSTRUMENTS; NONMETALS; RADIATION DETECTORS; RARE GASES; VARIATIONS