Efficiency calibration of HPGe detector and covariance analysis
- 1. Department of Physics, Bangalore University, Bengaluru (India)
- 2. Reactor Physics Division, Reactor Design and Development Group, Bhabha Atomic Research Centre, Mumbai (India)
- 3. Reactor Physics Division, Reactor Design and Development Group, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)
- 4. Department of Statistics, Manipal University, Manipal (India)
Description
Energy-efficiency calibration of the HPGe detector and corresponding covariance analysis may be considered as integral parts in the determination of nuclear cross-section. In the present work, gamma spectroscopy measurement using co-axial n-type HPGe detector (DSG-German) coupled to a PC-based 16-K channel Multiport-II MCA (Canberra), efficiency calibration and corresponding covariance analysis have been investigated. The resolution of the detector system during counting was 0.14% at 1332.5keV γ-line of 60Co. The standard calibration sources considered for the analysis are 133Ba, 22Na, 137Cs and 60Co. The covariance information obtained for the efficiencies of the HPGe detector with respect to γ-lines of standard calibration sources is further employed in the covariance analysis of efficiencies of the HPGe detector with respect to characteristic γ- lines of the reaction product 116mIn. (author)
Additional details
Publishing Information
- Publisher
- Bangalore University
- Imprint Place
- Bengaluru (India)
- Imprint Title
- Proceedings of the national conference on radiation physics
- Imprint Pagination
- 316 p.
- Journal Page Range
- p. 166-169
Conference
- Title
- national conference on radiation physics
- Acronym
- NCRP-2017
- Dates
- 23-24 Nov 2017
- Place
- Bengaluru (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 49021960
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BARIUM 133; CESIUM 137; COBALT 60; ENERGY EFFICIENCY; HIGH-PURITY GE DETECTORS; INDIUM 116
- Descriptors DEC
- ALKALINE EARTH ISOTOPES; BARIUM ISOTOPES; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; COBALT ISOTOPES; DAYS LIVING RADIOISOTOPES; EFFICIENCY; ELECTRON CAPTURE RADIOISOTOPES; EVEN-ODD NUCLEI; GE SEMICONDUCTOR DETECTORS; INDIUM ISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; SECONDS LIVING RADIOISOTOPES; SEMICONDUCTOR DETECTORS; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 2 refs., 2 figs., 3 tabs.