Published September 1982 | Version v1
Journal article

Image techniques for multiplanar computed tomography

  • 1. Duke Univ. Medical Center, Durham, NC

Description

Hardware and software options of the GE 8800 CT scanner were analyzed with respect to their effect on image quality for multiplanar images. Phantom studies were undertaken to demonstrate the effect of collimation, thick pixel reconstruction, and interpolation of both high- and low-contrast multiplanar images. Noise and spatial resolution were measured. Thick pixel reconstruction was found to be most useful in aiding in the delineation of low-contrast lesion boundaries. In addition, this option permits use of lower techniques, thus speeding data acquisition and reducing patient dose. Clinical examples are included

Additional details

Publishing Information

Journal Title
Radiology
Journal Volume
144
Journal Issue
4
Series
Radiology.
Journal Page Range
829-834
ISSN
0033-8419

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14742301
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; COMPUTERIZED TOMOGRAPHY; IMAGE PROCESSING; IMAGE SCANNERS; PATIENTS; PHANTOMS; RADIATION DOSES; SPATIAL RESOLUTION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; MOCKUP; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY