Published 2018 | Version v1
Book

Influence of various parameters on alpha beta discrimination using HIDEX 300SL liquid scintillation counter

  • 1. Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai (India)

Description

One of the vital features of HIDEX 300 SL Liquid Scintillation Counter (LSC) is α/β discrimination. For efficient α/β separation, various counting parameters such as Pulse length Index (PLI), Alpha Delay Time (ADT) and Alpha Tail Offset (ATO), need to be optimized HIDEX LSC analyzes the pulse after 'alpha delay time' (ADT) so that it collects mainly the delayed component (Application note: HIDEX, 2012). The purpose of this study was to standardize the calibration procedure for simultaneous determination of α/β activities in a sample using HIDEX LSC. % deviation in activities was studied as a function of PLI, ADT and ATO. Effect of sample quenching on PLI settings was also investigated

Part of:
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts

Additional details

Publishing Information

Publisher
Indian Association for Radiation Protection
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the thirty-third IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment: book of abstracts
Imprint Pagination
314 p.
Journal Page Range
p. 235

Conference

Title
33. IARP international conference on developments towards improvement of radiological surveillance at nuclear facilities and environment
Acronym
IARPIC-2018
Dates
16-20 Jan 2018
Place
Mumbai (India)

Optional Information

Notes
1 ref., 2 figs.