Published 1986 | Version v1
Report Open

Quantitative microanalysis of hafnium - zirconium system by X-ray fluorescence

Description

X-ray fluorescence technique has been used for the analysis of Hf in the presence of Zr by developing a method. In this method the spectral interference of Hf lines by Zr is eliminated completely and the Hf detection is accomplished using the most efficient Li line of its L-series. The principle of the method is based on the extinction properties of crystals for some orders of reflection. Ge(III) is used as the analyzing crystal. This method can be used accurately to detect Hf in any concentration of Zr at least from about 20 ppm to 100%. Also no information about the expected range of the analyte sample, is required in advance. (authors)

Availability note (English)

MF available from INIS under the Report Number.

Files

18039913.pdf

Files (678.2 kB)

Name Size Download all
md5:fdcae9ee1a9a9d2cb812106ab86d60a3
678.2 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
15 p.
Report number
PINSTECH/NMD--115

Optional Information

Notes
10 refs.; 5 figs.