Published November 14, 1983 | Version v1
Journal article

Reproducibility of ejection fraction measurements by gated equilibrium blood pool scintigraphy

  • 1. Vrije Univ., Amsterdam (Netherlands). Dept. of Nuclear Medicine
  • 2. Vrije Univ., Amsterdam (Netherlands). Dept. of Medical Statistics
  • 3. Vrije Univ., Amsterdam (Netherlands). Dept. of Cardiology
  • 4. Vrije Univ., Amsterdam (Netherlands). Dept. of Medical Informatics

Description

The reproducibility of ejection fraction measurements has been studied using gated equilibrium blood pool scintigraphy. The use of appropriate statistical tests is proposed and commented upon. The intra-observer variability for our group of patients has a standard deviation of 6.4%, the interobserver variability of 3.2% and sequential studies done on the same and different days give standard deviations (due to ''time'' alone) of 1.0% and 1.9% respectively. Different factors and sources involved in variability are mentioned. Variability values reported in the literature are discussed. (orig.)

Additional details

Publishing Information

Journal Title
Eur. J. Nucl. Med.
Journal Volume
8
Journal Issue
11
Series
CODEN: EJNMD.;Eur. J. Nucl. Med.
Journal Page Range
467-470
ISSN
0340-6997