Automatic radiographic quantification of hand osteoarthritis; accuracy and sensitivity to change in joint space width in a phantom and cadaver study
Creators
- 1. Leiden University Medical Center, Department of Radiology, Leiden (Netherlands)
- 2. Leiden University Medical Center, Division of Image Processing, Leiden (Netherlands)
- 3. Leiden University Medical Center, Department of Orthopaedics, Leiden (Netherlands)
- 4. Leiden University Medical Center, Department of Rheumatology, Leiden (Netherlands)
Description
To validate a newly developed quantification method that automatically detects and quantifies the joint space width (JSW) in hand radiographs. Repeatability, accuracy and sensitivity to changes in JSW were determined. The influence of joint location and joint shape on the measurements was tested. A mechanical micrometer set-up was developed to define and adjust the true JSW in an acrylic phantom joint and in human cadaver-derived phalangeal joints. Radiographic measurements of the JSW were compared to the true JSW. Repeatability, systematic error (accuracy) and sensitivity (defined as the smallest detectable difference (SDD)) were determined. The influence of joint position on the JSW measurement was assessed by varying the location of the acrylic phantom on the X-ray detector with respect to the X-ray beam and the influence of joint shape was determined by using morphologically different human cadaver joints. The mean systematic error was 0.052 mm in the phantom joint and 0.210 mm in the cadaver experiment. In the phantom experiments, the repeatability was high (SDD=0.028 mm), but differed slightly between joint locations (p=0.046), and a change in JSW of 0.037 mm could be detected. Dependent of the joint shape in the cadaver hand, a change in JSW between 0.018 and 0.047 mm could be detected. The automatic quantification method is sensitive to small changes in JSW. Considering the published data of JSW decline in the normal and osteoarthritic population, the first signs of OA progression with this method can be detected within 1 or 2 years. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00256-011-1110-xAdditional details
Identifiers
Publishing Information
- Journal Title
- Skeletal Radiology
- Journal Volume
- 41
- Journal Issue
- 1
- Journal Page Range
- p. 41-49
- ISSN
- 0364-2348
- CODEN
- SKRADI
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 43043898
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BIOMEDICAL RADIOGRAPHY; BONE JOINTS; CLINICAL TRIALS; DISTANCE; ERRORS; HANDS; PATHOLOGY; PHANTOMS; POLYACRYLATES; RHEUMATIC DISEASES; SENSITIVITY; SKELETAL DISEASES; SPATIAL RESOLUTION; SPECIFICITY; VALIDATION; X-RAY RADIOGRAPHY
- Descriptors DEC
- ARMS; BODY; DIAGNOSTIC TECHNIQUES; DISEASES; ESTERS; INDUSTRIAL RADIOGRAPHY; LIMBS; MATERIALS TESTING; MEDICINE; MOCKUP; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; ORGANIC POLYMERS; ORGANS; POLYMERS; POLYVINYLS; RADIOLOGY; RESOLUTION; SKELETON; STRUCTURAL MODELS; TESTING