Towards an optimum design of a P-MOS radiation detector for use in high-energy medical photon beams and neutron facilities: Analysis of activation materials

Creators

  • 1. Dept. of Radiography, City Univ., Rutland Place, Charterhouse Square, London EC1M 6PA (United Kingdom)

Description

The behaviour of packaged and un-packaged ESAPMOS4 RadFET radiation detectors (NMRC Cork, Ireland) was investigated when used in the mixed photon and neutron environment of a medical linear accelerator operating above the nucleon separation energy and in a 14 MeV neutron field provided by a D-T generator. Within the uncertainty of the experimental set-up (4% at 95% confidence level) the un-packaged device was found to have essentially zero activation dose-burden whereas the packaged device exhibits a considerable degree of post irradiation absorbed dose due to deactivation radiation. (authors)