SU-E-J-259: Quantification of Rotational Localization Offset in Radiation Therapy

  • 1. Rhode Island Hospital / Warren Alpert Medical, Providence, RI (United States)

Description

Purpose: To verify the necessity of 6D localization detection and correction in radiation therapy. Methods: An anthropomorphic head and neck phantom was used to test the BrainLab ExacTrac x-ray imaging system. After initial positioning, both ExacTrac and the on-board kV CBCT were used to detect the offset at the same position, using both manual and automatic registration algorithms. Then 6D offset including rotational errors up to 5 degree were introduced. Both ExacTrac and CBCT were used to correct the offsets and the residual errors were evaluated. Finally, 6D offset detected by ExacTrac for a C-Spine patient was reported. Results: The differences in 3D offset detected by ExacTrac and CBCT were 1.5 ± 1.2(Lateral), 2.7 ± 2.7(Vertical), and 4.0 ± 6.3(longitudinal) mm with manual registration while the corresponding differences decreased to 0.6 ± 0.3, 1.0 ± 0.3, and 0.3 ± 0.3 when automatic registration were used. CBCT corrected the translational offset to within 0.5 mm but the rotational errors remained and detected by the ExacTrac system (Yaw=2.1, Roll=1.1, Pitch=1.4 degree). When similar offset was introduced and corrected using ExacTrac, the residual error detected by both CBCT and ExacTrac were within 0.5 mm / 0.5 degree. The average offset from the 112 ExacTrac x-ray corrections for the C-Spine patient was 0.6 ± 1.6 (lateral), 5.4 ± 8(vertical), 1.6±1.1(longitudinal) mm, and 0.7 ± 0.6 (pitch), 0.7 ± 0.4(roll), 1.2 ± 0.7 (yaw) degree. Larger rotational errors, with a maximum of 2.7 degree (corresponds to about 1.5 to 4.5 mm offset for a POI 10 to 30 cm away from the isocenter), were observed when couch rotational were involved. Conclusion: Rotational errors are common in patient localization, especially when couch rotation is involved. Both appropriate imaging system and 6D robotic couch are necessary to detect and correct the rotational localization errors

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
6
Journal Page Range
p. 217
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46046105
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S60: APPLIED LIFE SCIENCES;
Descriptors DEI
ALGORITHMS; CORRECTIONS; ERRORS; HEAD; NECK; PATIENTS; PHANTOMS; PITCHES; RADIOTHERAPY; VERTEBRAE
Descriptors DEC
BODY; MATHEMATICAL LOGIC; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; ORGANS; OTHER ORGANIC COMPOUNDS; RADIOLOGY; SKELETON; STRUCTURAL MODELS; THERAPY

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine