Published 1981
| Version v1
Conference paper
Comparison of LKB Wallac-Multigamma 1260 multidetector instrument with other instruments for gamma sample measurement
Creators
- 1. Palackeho Univ., Olomouc (Czechoslovakia). Prirodovedecka Fakulta
- 2. Fakultni Nemocnice, Olomouc (Czechoslovakia). Oddeleni Nuklearni Mediciny
Description
The Multigamma 1260 consists of 12 well-type NaI(Tl) scintillation crystals shielded with a 10 mm thick lead block. It is microprocessor controlled. Its energy range is 10 to 800 keV. The instrument allows measuring 660 samples per hour at a measurement time of 1 min. Efficiency for 125I detection is minimally 75%. A comparison was made with counters Searle 4227 (100 samples) and Searle 1185 (300 samples) whose operating schedule is much slower. Long-term and short-term stabilities of the instruments and its volume dependence for 125I were tested. (H.S.)
Additional details
Additional titles
- Original title (Czech)
- Vysledky porovnani vicedetektoroveho pristroje LKB Wallac-Multigamma 1260 s jinymi aparaturami pro mereni vzorku gama
Publishing Information
- Imprint Title
- Nucleonic instruments and their use in the CSSR
- Imprint Pagination
- 280 p.
- Journal Page Range
- p. 61-66.
- Report number
- INIS-mf--8584
Conference
- Title
- Conference on nucleonic instruments and their use in the CSSR.
- Dates
- 18-22 May 1981.
- Place
- Pribram (Czechoslovakia).
INIS
- Country of Publication
- Serbia and Montenegro
- Country of Input or Organization
- Serbia and Montenegro
- INIS RN
- 14780177
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- EFFICIENCY; GAMMA SPECTROMETERS; NAI DETECTORS; SCINTILLATION COUNTING; SPECIFICATIONS; STABILITY
- Descriptors DEC
- COUNTING TECHNIQUES; MEASURING INSTRUMENTS; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTROMETERS
Optional Information
- Notes
- Imprint:Pristroje jaderne techniky a jejich vyuziti v CSSR.