Published 1981 | Version v1
Conference paper

Comparison of LKB Wallac-Multigamma 1260 multidetector instrument with other instruments for gamma sample measurement

  • 1. Palackeho Univ., Olomouc (Czechoslovakia). Prirodovedecka Fakulta
  • 2. Fakultni Nemocnice, Olomouc (Czechoslovakia). Oddeleni Nuklearni Mediciny

Description

The Multigamma 1260 consists of 12 well-type NaI(Tl) scintillation crystals shielded with a 10 mm thick lead block. It is microprocessor controlled. Its energy range is 10 to 800 keV. The instrument allows measuring 660 samples per hour at a measurement time of 1 min. Efficiency for 125I detection is minimally 75%. A comparison was made with counters Searle 4227 (100 samples) and Searle 1185 (300 samples) whose operating schedule is much slower. Long-term and short-term stabilities of the instruments and its volume dependence for 125I were tested. (H.S.)

Additional details

Additional titles

Original title (Czech)
Vysledky porovnani vicedetektoroveho pristroje LKB Wallac-Multigamma 1260 s jinymi aparaturami pro mereni vzorku gama

Publishing Information

Imprint Title
Nucleonic instruments and their use in the CSSR
Imprint Pagination
280 p.
Journal Page Range
p. 61-66.
Report number
INIS-mf--8584

Conference

Title
Conference on nucleonic instruments and their use in the CSSR.
Dates
18-22 May 1981.
Place
Pribram (Czechoslovakia).

INIS

Country of Publication
Serbia and Montenegro
Country of Input or Organization
Serbia and Montenegro
INIS RN
14780177
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
EFFICIENCY; GAMMA SPECTROMETERS; NAI DETECTORS; SCINTILLATION COUNTING; SPECIFICATIONS; STABILITY
Descriptors DEC
COUNTING TECHNIQUES; MEASURING INSTRUMENTS; RADIATION DETECTORS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTROMETERS

Optional Information

Notes
Imprint:Pristroje jaderne techniky a jejich vyuziti v CSSR.