Published January 2002
| Version v1
Journal article
Development of simulated-gas standards
Creators
Description
Multigamma simulated-gas standards were developed for monitoring the surrounding air at nuclear power plants by several commercial firms and laboratories. The main advantages of our simulated-gas standard consist of the easy preparation, gravimetrically spiking the matrix with a standardised multigamma radioactive solution, and that it does not require to be tested against actual gas standards. Self-attenuation correction factors for low γ-ray energies were calculated. Original studies about the stability of the adherence of the dried radioactive residue to the matrix are presented and discussed
Additional details
Identifiers
- PII
- S0969804301001658;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 56
- Journal Issue
- 1-2
- Journal Page Range
- p. 47-50
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 33047104
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- CALIBRATION STANDARDS; GAMMA SOURCES; GAMMA SPECTROSCOPY; GASES; RADIATION MONITORING; SAMPLE PREPARATION; SIMULATION
- Descriptors DEC
- FLUIDS; MONITORING; RADIATION SOURCES; SPECTROSCOPY; STANDARDS
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.