Published 1994 | Version v1
Journal article

Indoor radon daughter contamination monitoring: the absolute efficiency of CR-39 taking into account the plateout effect and environmental conditions

  • 1. Universidade Estadual de Campinas, SP (Brazil). Inst. de Fisica
  • 2. Univ. Federal de Mato Grasso, Cuiaba, MT (Brazil)

Description

In this work it is shown that CR-39 (Supergrade PM355, Pershore Mouldings Ltd, England) can be used as an absolute detector in 222Rn and radon daughters (RD) monitoring by employing it as an α spectrometer. By adopting the technique described here, one can measure separately: (a) the 222Rn activity, (b) the RD α activity in the air, and (c) the RD α activity plated out on the detector surface. The calibration factors measured for these three items were found to be weakly dependent on environmental conditions. The relationship between the track density and the radon and daughters α activity taking into account the influence of the plateout phenomenon and environmental conditions is discussed. Possible applications of the experimental results on subjects related to dose calculation, like the mean RD deposition rates indoors, are also discussed. (Author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
51
Journal Issue
4
Journal Page Range
p. 283-296.
ISSN
0144-8420
CODEN
RPDODE

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
25050159
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
DAUGHTER PRODUCTS; DIELECTRIC TRACK DETECTORS; EFFICIENCY; INDOOR AIR CONTAMINATION; RADON
Descriptors DEC
CONTAMINATION; ELEMENTS; ISOTOPES; MEASURING INSTRUMENTS; NONMETALS; RADIATION DETECTORS; RARE GASES