'Accord 12' multichannel analyzer for simultaneous recording deuterium and hydrogen atom energy spectra
- 1. AN SSSR, Leningrad (Russian Federation). Fiziko-Tekhnicheskij Inst.
Description
A twelve-channel analyzer allowing one to simultaneously record the energy distribution and absolute fluxes of hydrogen and deuterium atoms emitted by high-temperature plasma is described. The analyser dynamic range i.e. the ratio of energies measured simultaneously by the edge device channels makes up ∼ 7. The minimum energy of the particles recorded is 200 eV the maximum is 70 keV. The energy channel width is 9-30%. Channel electron multipliers VEM-6 are used as detectors. The analyzer is calibrated on hydrogen and deuterium beams. Calibration is conducted under single atom collisions with nitrogen molecules under stationary admission to a stripping chamber. Atom recording efficiency falls within 10-5 (deuterium at 200 eV) - 2x10-1 (hydrogen at 70 keV) limits
Additional details
Additional titles
- Original title (Russian)
- Многоканальный анализатор для одновременной регистрации энергетических спектров атомов водорода и дейтерия 'Акорд-12ь
Publishing Information
- Journal Title
- Zhurnal Tekhnicheskoj Fiziki
- Journal Volume
- 62
- Journal Issue
- 2
- Journal Page Range
- p. 157-163.
- ISSN
- 0044-4642
- CODEN
- ZTEFA3
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- Russian Federation
- INIS RN
- 24060412
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; DEUTERIUM; EFFICIENCY; ELECTRON MULTIPLIERS; ENERGY SPECTRA; HYDROGEN; ION DETECTION; MASS SPECTROMETERS; MULTI-CHANNEL ANALYZERS; PLASMA DIAGNOSTICS; SENSITIVITY; SPECIFICATIONS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; ELECTRON TUBES; ELECTRONIC EQUIPMENT; ELEMENTS; EQUIPMENT; HYDROGEN ISOTOPES; ISOTOPES; LIGHT NUCLEI; MEASURING INSTRUMENTS; NONMETALS; NUCLEI; ODD-ODD NUCLEI; PULSE ANALYZERS; RADIATION DETECTION; SPECTRA; SPECTROMETERS; STABLE ISOTOPES