Methodology for the construction of a physical phantom for quality control of images in digital radiography
Creators
- 1. Instituto Federal de Educacao, Ciencia e Tecnologia de Pernambuco (IFFPE), Recife, PE (Brazil)
- 2. Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil)
- 3. Universidade Federal de Pernambuco (UFPE), Recife, PE (Brazil)
Description
The advancement of technology in recent years has provided the production of increasingly sophisticated devices, aiming to acquire medical images with high technical level and also facilitate the operational readiness of the equipment. In order to ensure the most accurate diagnosis with minimum dose without exposing patients to obtain data and verify the performance of a radiographic system for quality control purposes we use the so-called phantoms. Phantoms are physical or computational models used to simulate the transport of ionizing radiation, their interactions in the tissues of the human body and evaluate the deposition of energy. Besides, they are made from materials with behavior similar to human tissues when exposed to ionizing radiation - the so-called tissue-equivalent materials. This paper describes the construction of a physical phantom that allows the execution of the main acceptance tests of the quality control protocols in digital radiography
Files
45068147.pdf
Files
(321.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:111bc36bbf8ce989f44a2534473322ce
|
321.2 kB | Preview Download |
Additional details
Additional titles
- Original title (Portuguese)
- Metodologia para a construcao de um fantoma fisico para controle de qualidade de imagens em radiografia digital
Publishing Information
- Imprint Pagination
- 9 p.
- Report number
- INIS-BR--13943
Conference
- Title
- International nuclear atlantic conference
- Acronym
- INAC 2013
- Dates
- 24-29 Nov 2013
- Place
- Recife, PE (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 45068147
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DIGITAL SYSTEMS; FABRICATION; IMAGES; IONIZING RADIATIONS; PHANTOMS; QUALITY CONTROL; SIMULATION; X-RAY EQUIPMENT
- Descriptors DEC
- CONTROL; DIAGNOSTIC TECHNIQUES; EQUIPMENT; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; STRUCTURAL MODELS
Optional Information
- Notes
- 13 refs., 5 figs., 1 tab.