Published April 2007
| Version v1
Journal article
TLD international intercomparision of secondary dosimetry standard for radiotherapy level in 2005
Creators
Description
Objective: To check the consistency of SSDL radiotherapy dose level standard and international standard, SSDL in our institute participated in TLD international intercomparision organized by IAEA/WHO. Methods: SSDL irradiated the TLDs posted by IAEA and calculated the dose to be delivered to the TLDs, then returned the TLDs and results to IAEA dosimetry laboratory. IAEA gave the diviation of intercomparision. Results: The deviation for 60Co is -1.4%, and the deviation for high energy X rays is - 1.8 %. Conclusions: The requirement of deviation between IAEA and SSDL should be no more than ±3.5%. Therefore the results of this intercomparison are considered satisfactory. (authors)
Additional details
Publishing Information
- Journal Title
- Chinese Journal of Radiological Medicine and Protection
- Journal Volume
- 27
- Journal Issue
- 2
- Journal Page Range
- p. 181-182
- ISSN
- 0254-5098
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 40060261
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COBALT 60; DOSIMETRY; IAEA; INTERLABORATORY COMPARISONS; IRRADIATION; RADIATION DOSES; RADIOTHERAPY; SSDL; STANDARDS; WHO; X RADIATION
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; DOSES; ELECTROMAGNETIC RADIATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; INTERNATIONAL ORGANIZATIONS; IONIZING RADIATIONS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEDICINE; MINUTES LIVING RADIOISOTOPES; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; RADIATIONS; RADIOISOTOPES; RADIOLOGY; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 2 tabs., 5 refs.