Published September 2000 | Version v1
Journal article

Hyperboloid mass spectrometer with a monopolar ion trap

  • 1. Ryazanskaya Gosudarstvennaya Radiotekhnicheskaya Akademiya, Ryazan' (Russian Federation)

Description

Paper describes a hyperboloid mass-spectrometer with three-dimensional ion trap base analyzer. Potential distribution in monopolar trap was numerically simulated and the effect of ΔU(z, r) distortion function on the paths of the charged particles was studied. Mass peaks were estimated for the analyzed various operation conditions. The experimental mass-spectrometer with monopolar ion trap was studied. The resolution equal to 1.2 x 103 was reached. Application of monopolar ways for mass-selective classification of ions in nonlinear HF-fields enables to simplify design of their electrode system, to solve efficiently the problem of charge particle admission and remove in case of high resolution of hyperboloid analyzers

Abstract (Russian)

Описан гиперболоидный масс-спектрометр с анализатором на основе трехмерной ионной ловушки. Проведено численное моделирование распределения потенциала в монополярной ловушке и исследовано влияние функции искажений ΔU(z, r) на траектории заряженных частиц. Рассчитаны массовые пики для различных режимов работы анализатора. Исследован экспериментальный масс-спектрометр с монополярной ионной ловушкой. Достигнута разрешающая способность 1,2 х 103. Применение монополярных способов масс-селективной сортировки ионов в нелинейных ВЧ-полях позволяет при высоком разрешении гиперболоидных анализаторов упростить конструкцию их электродной системы, эффективно решить проблему ввода и вывода заряженных частиц

Additional details

Additional titles

Original title (Russian)
Гиперболоидный масс-спектрометр с монополярной ионной ловушкой

Publishing Information

Journal Title
Pribory i Tekhnika Ehksperimenta
Journal Issue
no.5
Journal Page Range
p. 59-63
ISSN
0032-8162
CODEN
PRTEAJ

Optional Information

Notes
6 refs., 7 figs.