Published October 2012 | Version v1
Journal article

A photodiode-based neutral particle bolometer for characterizing charge-exchanged fast-ion behavior

  • 1. Tri Alpha Energy, Inc., Rancho Santa Margarita, California 92688 (United States)
  • 2. University of California-Irvine, Irvine, California 92697 (United States)

Description

A neutral particle bolometer (NPB) has been designed and implemented on Tri Alpha Energy's C-2 device in order to spatially and temporally resolve the charge-exchange losses of fast-ion populations originating from neutral beam injection into field-reversed configuration plasmas. This instrument employs a silicon photodiode as the detection device with an integrated tungsten filter coating to reduce sensitivity to light radiation. Here we discuss the technical aspects and calibration of the NPB, and report typical NPB measurement results of wall recycling effects on fast-ion losses.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
83
Journal Issue
10
Journal Page Range
p. 10D713-10D713.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2012 American Institute of Physics