Capacitively coupled pickup in MCP-based photodetectors using a conductive metallic anode
- 1. Enrico Fermi Institute, University of Chicago, 5640 S Ellis Ave, Chicago, IL 60637 (United States)
- 2. Incom, Inc., 294 SouthBridge Rd, Charlton, Massachusetts 01507 (United States)
Description
We have designed and tested a robust 20×20 cm2 thin metal film internal anode capacitively coupled to an external array of signal pads or micro-strips for use in fast microchannel plate photodetectors. The internal anode, in this case a 10 nm-thick NiCr film deposited on a 96% pure Al2O3 3 mm-thick ceramic plate and connected to HV ground, provides the return path for the electron cascade charge. The multi-channel pickup array consists of a printed-circuit card or glass plate with metal signal pickups on one side and the signal ground plane on the other. The pickup can be put in close proximity to the bottom outer surface of the sealed photodetector, with no electrical connections through the photodetector hermetic vacuum package other than a single ground connection to the internal anode. Two pickup patterns were tested using a small commercial MCP-PMT as the signal source: 1) parallel 50 Ω 25-cm-long micro-strips with an analog bandwidth of 1.5 GHz, and 2) a 20×20 cm2 array of 2-dimensional square 'pads' with sides of 1.27 cm or 2.54 cm. The rise-time of the fast input pulse is maintained for both pickup patterns. For the pad pattern, we observe 80% of the directly coupled amplitude. For the strip pattern we measure 34% of the directly coupled amplitude on the central strip of a broadened signal. The physical decoupling of the photodetector from the pickup pattern allows easy customization for different applications while maintaining high analog bandwidth.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2016.12.008Additional details
Identifiers
- DOI
- 10.1016/j.nima.2016.12.008;
- arXiv
- arXiv:1610.01434v1;
- PII
- S0168-9002(16)31251-7;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 846
- Journal Page Range
- p. 75-80
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48082708
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALUMINIUM OXIDES; AMPLITUDES; ANODES; CERAMICS; ELECTRONS; GHZ RANGE 01-100; METALS; MICROCHANNEL ELECTRON MULTIPLIERS; PHOTODETECTORS; PHOTOMULTIPLIERS; PLATES; PRINTED CIRCUITS; PULSE RISE TIME; SIGNALS; SURFACES; THIN FILMS
- Descriptors DEC
- ALUMINIUM COMPOUNDS; CHALCOGENIDES; ELECTRODES; ELECTRON MULTIPLIERS; ELECTRON TUBES; ELECTRONIC CIRCUITS; ELEMENTARY PARTICLES; ELEMENTS; FERMIONS; FILMS; FREQUENCY RANGE; GHZ RANGE; LEPTONS; OXIDES; OXYGEN COMPOUNDS; PHOTOTUBES; TIMING PROPERTIES
Optional Information
- Copyright
- Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.