Published April 1989
| Version v1
Journal article
On the use of automatic pipets in volume calibration of accountability tanks
Description
The applicability of automatic pipets for dispensing known volumes of water in the volume calibration of accountability tanks of tens-of-liters capacity is demonstrated. Six pipet of capacity in the approximate range 495 to 716 milliliters were calibrated using gravimetric techniques. The relative standard deviation ranged from 0.0015% to 0.0074%. The pipets were used successfully in the volume calibration of accountability tanks of tens-of-liters capacity in the Savannah River Plant
Additional details
Publishing Information
- Journal Title
- JNMM
- Journal Volume
- 17
- Journal Issue
- 3
- Series
- JNMM.
- Journal Page Range
- 32-33
- ISSN
- 0893-6188
- CODEN
- JNMME
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21003620
- Subject category
- S98: NUCLEAR DISARMAMENT, SAFEGUARDS AND PHYSICAL PROTECTION; S42: ENGINEERING;
- Descriptors DEI
- AUTOMATION; CALIBRATION; GRAVIMETRY; SAVANNAH RIVER PLANT; TANKS
- Descriptors DEC
- CONTAINERS; NATIONAL ORGANIZATIONS; US AEC; US DOE; US ERDA; US ORGANIZATIONS