Published August 2014 | Version v1
Journal article

Poster — Thur Eve — 43: Monte Carlo Modeling of Flattening Filter Free Beams and Studies of Relative Output Factors

  • 1. Grand River Regional Cancer Centre, Grand River Hospital, Kitchener, ON (Canada)

Description

Flattening filter free (FFF) beams have been adopted by many clinics and used for patient treatment. However, compared to the traditional flattened beams, we have limited knowledge of FFF beams. In this study, we successfully modeled the 6 MV FFF beam for Varian TrueBeam accelerator with the Monte Carlo (MC) method. Both the percentage depth dose and profiles match well to the Golden Beam Data (GBD) from Varian. MC simulations were then performed to predict the relative output factors. The in-water output ratio, Scp, was simulated in water phantom and data obtained agrees well with GBD. The in-air output ratio, Sc, was obtained by analyzing the phase space placed at isocenter, in air, and computing the ratio of water Kerma rates for different field sizes. The phantom scattering factor, Sp, can then be obtained from the traditional way of taking the ratio of Scp and Sc. We also simulated Sp using a recently proposed method based on only the primary beam dose delivery in water phantom. Because there is no concern of lateral electronic disequilibrium, this method is more suitable for small fields. The results from both methods agree well with each other. The flattened 6 MV beam was simulated and compared to 6 MV FFF. The comparison confirms that 6 MV FFF has less scattering from the Linac head and less phantom scattering contribution to the central axis dose, which will be helpful for improving accuracy in beam modeling and dose calculation in treatment planning systems

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
8
Journal Page Range
p. 16
ISSN
0094-2405
CODEN
MPHYA6

INIS

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine