Published January 2015 | Version v1
Journal article

Dosimetry of consecutive linear arranged 125I seeds

  • 1. Department of Interventional Radiology, Zhongshan Hospital, Fudan University, Shanghai (China)

Description

Objective: To investigate the radiation distribution of consecutive linear arranged 125I seeds in vitro. Methods: The in vitro 3-D radiation distribution was measured by radiophoto-luminescent glass detectors and imaging plate (IP) using a SM-2 model. The results were then compared with the data obtained from the software calculation. Results: Strong correlation existed between the software calculation and data obtained from radiophoto-luminescent glass detector measurements (r > 0.99, P = 0.000). No statistically significant difference was found between the two methods except at R = 0.5 cm area (P < 0.05). A proportion about 22.8% self-absorption within 2 cm was observed when IP measurement was used. Conclusion: The isodose lines of consecutive linear arranged 125I seeds conform to tubular structure, which provides theoretical basis for its application in intraluminal brachytherapy. (authors)

Additional details

Publishing Information

Journal Title
Journal of Interventional Radiology
Journal Volume
24
Journal Issue
1
Journal Page Range
p. 59-63
ISSN
1008-794X

Optional Information

Notes
3 figs., 29 refs.; http://dx.doi.org/10.3969/j.issn.1008-794X.2015.01.014