Published May 1, 2018 | Version v1
Journal article

Inductance analyzer based on auto-balanced circuit for precision measurement of fluxgate impedance

  • 1. Institut für Elektrische Messtechnik und Grundlagen der Elektrotechnik, TU Braunschweig, Hans-Sommer-Str. 66 D-38106 Braunschweig (Germany)

Description

An instrument for fluxgate sensor impedance measurement based on an auto-balanced circuit has been designed and characterized. The circuit design is adjusted to comply with the fluxgate sensor characteristics which are low impedance and highly saturable core with very high permeability. The system utilizes a NI-DAQ card and LabVIEW to process the signal acquisition and evaluation. Some fixed reference resistances are employed for system calibration using linear regression. A multimeter HP 34401A and impedance analyzer Agilent 4294A are used as calibrator and validator for the resistance and inductance measurements. Here, we realized a fluxgate analyzer instrument based on auto-balanced circuit, which measures the resistance and inductance of the device under test with a small error and much lower excitation current to avoid core saturation compared to the used calibrator. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6501/aaae97

Additional details

Identifiers

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
29
Journal Issue
5
Journal Page Range
[9 p.]
ISSN
0957-0233
CODEN
MSTCEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
51046693
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; CALIBRATION; CURRENTS; DESIGN; EXCITATION; IMPEDANCE; INDUCTANCE; PERMEABILITY; SATURATION; SENSORS; SIGNALS
Descriptors DEC
ELECTRICAL PROPERTIES; ENERGY-LEVEL TRANSITIONS; PHYSICAL PROPERTIES