Inductance analyzer based on auto-balanced circuit for precision measurement of fluxgate impedance
- 1. Institut für Elektrische Messtechnik und Grundlagen der Elektrotechnik, TU Braunschweig, Hans-Sommer-Str. 66 D-38106 Braunschweig (Germany)
Description
An instrument for fluxgate sensor impedance measurement based on an auto-balanced circuit has been designed and characterized. The circuit design is adjusted to comply with the fluxgate sensor characteristics which are low impedance and highly saturable core with very high permeability. The system utilizes a NI-DAQ card and LabVIEW to process the signal acquisition and evaluation. Some fixed reference resistances are employed for system calibration using linear regression. A multimeter HP 34401A and impedance analyzer Agilent 4294A are used as calibrator and validator for the resistance and inductance measurements. Here, we realized a fluxgate analyzer instrument based on auto-balanced circuit, which measures the resistance and inductance of the device under test with a small error and much lower excitation current to avoid core saturation compared to the used calibrator. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6501/aaae97Additional details
Identifiers
Publishing Information
- Journal Title
- Measurement Science and Technology
- Journal Volume
- 29
- Journal Issue
- 5
- Journal Page Range
- [9 p.]
- ISSN
- 0957-0233
- CODEN
- MSTCEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 51046693
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CALIBRATION; CURRENTS; DESIGN; EXCITATION; IMPEDANCE; INDUCTANCE; PERMEABILITY; SATURATION; SENSORS; SIGNALS
- Descriptors DEC
- ELECTRICAL PROPERTIES; ENERGY-LEVEL TRANSITIONS; PHYSICAL PROPERTIES