Published October 1986
| Version v1
Journal article
Evaluation of the response of a round hole scintillation camera collimator by the Fourier analysis method
Creators
- 1. Hospital Clinico Universitario Zaragoza (Spain). Seccion de Fisica
- 2. Universidad Autonoma de Barcelona (Spain). Facultad de Ciencias
Description
The Fourier analysis method was used to investigate the response of scintillation camera collimators with parallel holes. This method which takes into account the septal penetration was applied to the case of round hole collimators having a hexagonal distribution. Modulation transfer functions, MTF have been determined to verify the accuracy of the computed Fourier coefficients of the collimator function. Comparisons between the geometric and the penetrating plus geometric transfer function are shown for round and hexagonal holes. (author)
Additional details
Publishing Information
- Journal Title
- Nucl. Med. Commun.
- Journal Volume
- 7
- Journal Issue
- 10
- Series
- Nucl. Med. Commun.
- Journal Page Range
- 779-785
- ISSN
- 0143-3636
- CODEN
- NMCOD
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 18024103
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COLLIMATORS; EVALUATION; FOURIER ANALYSIS; GAMMA CAMERAS; RESPONSE FUNCTIONS
- Descriptors DEC
- CAMERAS; FUNCTIONS