Published September 20, 1986
| Version v1
Journal article
Delay-line readout for multicell detectors
Creators
- 1. Grenoble-1 Univ., 74 - Annecy (France). Lab. de Physique des Particules
Description
A simple delay-line readout method has been tested with a position-sensitive photomultiplier. Position accuracies of the reconstruction by means of a time-to-amplitude convertor are quite comparable to those obtained by a conventional centroid method using a number of analog-to-digital convertors. In this way, a simple and economic solution can be conceived for the readout of not only position-sensitive photomultipliers, but also arrays of different kinds of position detectors. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res., Sect. A
- Journal Volume
- 250
- Journal Issue
- 3
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 539-542
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 17089709
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DELAY CIRCUITS; PHOTOMULTIPLIERS; POSITION SENSITIVE DETECTORS; READOUT SYSTEMS; TIME-TO-AMPLITUDE CONVERTERS
- Descriptors DEC
- ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS; PHOTOTUBES; PULSE CONVERTERS; RADIATION DETECTORS