Published December 1996
| Version v1
Miscellaneous
Open
Energy calibration of the 3 MV tandem accelerator
Creators
Description
By this study, a graphite calorimeter has been developed to measure the average ion energy from electrostatic ion accelerator (3 MV Tandem Ion Accelerator of the Takasaki Ion Accelerators for Advanced Radiation Application). A graphite calorimeter was designed for measuring the ion beam energy in wide ion energy ranges (from 3 MeV to 10 MeV). By a couple of this study, the 1 H(15 N, α γ) 12 C resonant nuclear reactions (resonant energy 6.385 MeV and 13.355 MeV) has been applied for measuring the absolute ion beam energy serve as a comparative method. The obtained results were compared in terms of resolution and accuracy. Close agreement was found for both methods, the deviation was observed within ± 3%
Files
29044579.pdf
Files
(316.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:7315509a8a22b141a09342b951224171
|
316.0 kB | Preview Download |
Additional details
Publishing Information
- Imprint Title
- Proceedings of the 6th nuclear science and technology conference
- Imprint Pagination
- 761 p.
- Journal Page Range
- [16 p.]
- Report number
- INIS-TH--006
Conference
- Title
- 6. Conference on Nuclear Science and Technology
- Dates
- 2-4 Dec 1996
- Place
- Bangkok (Thailand)
INIS
- Country of Publication
- Thailand
- Country of Input or Organization
- Thailand
- INIS RN
- 29044579
- Subject category
- S43: PARTICLE ACCELERATORS; S43: PARTICLE ACCELERATORS;
- Resource subtype / Literary indicator
- Conference, Numerical Data
- Descriptors DEI
- CALIBRATION; CALORIMETERS; EXPERIMENTAL DATA; TANDEM ELECTROSTATIC ACCELERATORS
- Descriptors DEC
- ACCELERATORS; DATA; ELECTROSTATIC ACCELERATORS; INFORMATION; MEASURING INSTRUMENTS; NUMERICAL DATA