Published July 1984 | Version v1
Journal article

Angled-collimator SPECT (A-SPECT): an improved approach to cranial single photon emission tomography

  • 1. Columbia Univ., New York, NY

Description

A special 300 slant-hole collimator has been developed for rotating camera single photon emission computerized tomography (SPECT) that allows the camera to remain close to the patient's head throughout a standard 3600 circular rotation. Compared with SPECT using parallel-hole collimation, angled-collimator SPECT (A-SPECT) yielded approximately a 30% increase in resolution without significant losses in sensitivity. The full width at half maximum in reconstructed transaxial images of Tc-99m line sources was 1.2 cm for the A-SPECT system compared with 1.6 cm for conventional SPECT, and A-SPECT yielded substantial improvements in modulation tranfer function for the line source. A-SPECT images of phantoms and of patients with paranasal sinus disease were consistently superior in resolution to those of conventional SPECT

Additional details

Publishing Information

Journal Title
J. Nucl. Med.
Journal Volume
25
Journal Issue
7
Series
J. Nucl. Med.
Journal Page Range
805-809
ISSN
0022-3123