Published March 1983 | Version v1
Journal article

Shaped, lead-loaded acrylic filters for patient exposure reduction and image-quality improvement

  • 1. Department of Diagnostic Radiology, Mayo Clinic and Foundation, Rochester, Minnesota

Description

Shaped filters that are constructed of lead-loaded acrylic material for use in patient radiography are discussed. Use of the filters will result in improved overall image quality with significant exposure reduction to the patient (approximately a 2X reduction in breast exposure and a 3X reduction in thyroid gland exposure). Detailed drawings of the shaped filters for scoliosis radiography, cervical spine radiography, and for long film changers in special procedures are provided. The use of the scoliosis filters is detailed and includes phantom and patient radiographs and dose reduction information

Additional details

Publishing Information

Journal Title
Radiology
Journal Volume
146
Journal Issue
3
Series
Radiology.
Journal Page Range
825-828
ISSN
0033-8419

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
15002228
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; FILTERS; IMAGES; PATIENTS; SHIELDING; SPINAL CORD; X-RAY EQUIPMENT
Descriptors DEC
CENTRAL NERVOUS SYSTEM; DIAGNOSTIC TECHNIQUES; EQUIPMENT; MEDICINE; NERVOUS SYSTEM