Published December 2015 | Version v1
Journal article

Note: Spectrometer with multichannel photon-counting detector for beam emission spectroscopy in magnetic fusion devices

  • 1. Novosibirsk State University, 630090 Novosibirsk (Russian Federation)
  • 2. Budker Institute of Nuclear Physics, 630090 Novosibirsk (Russian Federation)
  • 3. Novosibirsk State Technical University, 630073 Novosibirsk (Russian Federation)

Description

A spectrometer based on a linear array photomultiplier tube (PMT) has been developed and calibrated. A 0.635 m focal length Czerny-Turner monochromator combined with a coupling optics provides an image of a narrow 0.5 nm spectral range with a resolution of 0.015 nm/channel on a 32-anode PMT. The system aims at spectroscopy of Dα or Hα lines emitted by a diagnostic atomic beam in a plasma (primarily a motional Stark effect diagnostics). To record a low photon flux of ∼106 s−1 per channel with the time resolution of 100 μs, a pulse counting approach has been used. Wideband amplifiers scale single-electron pulses and transmit them to a digital data processing core hardwired in a programmable logic matrix. Calibrations have shown that the aberration-limited instrument function fits to a single detector channel of 1 mm width. Pilot results of passive measurements of Dα light emission from the plasma confined in a magnetic trap are presented

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
86
Journal Issue
12
Journal Page Range
p. 126109-126109.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2015 AIP Publishing LLC