Fidelity imaging for atomic force microscopy

  • 1. Nanodynamics Systems Laboratory, Department of Electrical and Computer Engineering, University of Minnesota, Minneapolis, Minnesota 55455 (United States)

Description

Atomic force microscopy is widely employed for imaging material at the nanoscale. However, real-time measures on image reliability are lacking in contemporary atomic force microscopy literature. In this article, we present a real-time technique that provides an image of fidelity for a high bandwidth dynamic mode imaging scheme. The fidelity images define channels that allow the user to have additional authority over the choice of decision threshold that facilitates where the emphasis is desired, on discovering most true features on the sample with the possible detection of high number of false features, or emphasizing minimizing instances of false detections. Simulation and experimental results demonstrate the effectiveness of fidelity imaging

Additional details

Identifiers

Publishing Information

Journal Title
Applied Physics Letters
Journal Volume
106
Journal Issue
1
Journal Page Range
p. 013113-013113.4
ISSN
0003-6951
CODEN
APPLAB

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46101370
Subject category
S77: NANOSCIENCE AND NANOTECHNOLOGY;
Descriptors DEI
ATOMIC FORCE MICROSCOPY; COMPUTERIZED SIMULATION; DETECTION; IMAGES; NANOSTRUCTURES; REAL TIME SYSTEMS
Descriptors DEC
MICROSCOPY; SIMULATION

Optional Information

Notes
(c) 2015 AIP Publishing LLC