Published February 15, 1982 | Version v1
Journal article

Calibration of low energy photon detectors for microbeam PIXE analysis

  • 1. Surrey Univ., Guildford (UK). Medical and Environmental Physics Group

Description

A low photon energy variable geometry gas flow proportional counter and a Si(Li) semiconductor detector are intercalibrated for use in proton induced X-ray emission analysis with a proton microbeam facility. From the ratio of photopeak counts, for the two detectors in situ, and theoretical calculations, the Si(Li) detector efficiency below approx. equal to2.3 keV is obtained. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res.
Journal Volume
193
Journal Issue
1/2
Series
Nucl. Instrum. Methods Phys. Res.
Journal Page Range
99-103
ISSN
0029-554X

Conference

Title
5. symposium on x- and gamma-ray sources and applications.
Dates
10 - 12 Jun 1981.
Place
Ann Arbor, MI, USA.