A High Sensitivity EPR Technique for Alanine Dosimetry (invited paper)
Creators
Description
Uncertainties of ± 5 mGy were achieved in the measurement of alanine dosemeters using optimised EPR parameters, instrumentation, spectral manipulation and subtraction techniques. Modulation amplitude and microwave power were adjusted to combine resonances of two neighbouring alanine signals. Instrumental variations were minimised by combining and subtracting pre- and post-measurement spectra of the empty EPR tube. A spectrum of the native signal of non-dosed alanine was generated from a single batch of dosemeters and subtracted from spectra of the irradiated dosemeters, also from the same batch. Field alignment was adjusted with the use of an in-cavity Mn++ standard. A constant rotation goniometer was used to eliminate anisotropies in the EPR tube and alanine samples. Finally, digital filters were applied to the resulting spectra. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 77
- Journal Issue
- 1-2
- Journal Page Range
- p. 43-49
- ISSN
- 0144-8420
- CODEN
- RPDODE
Conference
- Title
- International workshop on retrospective dosimetry (physical and biological aspects)
- Dates
- 14-17 Oct 1997
- Place
- Tianjin (China)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Argentina
- INIS RN
- 31039497
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALANINES; AMPLITUDES; DOSEMETERS; DOSIMETRY; ELECTRON SPIN RESONANCE; MICROWAVE RADIATION; MODULATION; RADIATION DOSES
- Descriptors DEC
- AMINO ACIDS; CARBOXYLIC ACIDS; DOSES; ELECTROMAGNETIC RADIATION; MAGNETIC RESONANCE; MEASURING INSTRUMENTS; ORGANIC ACIDS; ORGANIC COMPOUNDS; RADIATIONS; RESONANCE