Dosimetric aspects of a 3.3-MV linear accelerator
Creators
- 1. Department of Radiation Oncology, St. Barnabas Medical Center, Livingston, New Jersey 07039 (USA)
Description
Both the design considerations and the dosimetric properties of the Siemens Model 5800 linear accelerator are discussed. This unit is of such an energy (3.3 MV) as to imitate Cobalt-60 teletherapy depth doses. A linear relation of dmax to depth dose at low energies was found for various wave guides and targets. The energy of the unit can be characterized by its nominal accelerating potential of 2.70 MV, its d80 of 5.3 cm, its first half-value layer of 0.8 cm lead and the measured energy of the electron beam at 3.3 MeV. The following selected commissioning aspects are reported: central axis depth dose, relative output factors, beam profiles, wedge factors, virtual source position, back scatter factors, penumbra and build-up region
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 17
- Journal Issue
- 3
- Series
- Med. Phys.
- Journal Page Range
- 474-480
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21067896
- Subject category
- S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- BEAM PROFILES; BUILDUP; DEPTH DOSE DISTRIBUTIONS; LINEAR ACCELERATORS; PERFORMANCE; PHOTON BEAMS; SPECIFICATIONS; USES; X RADIATION; X-RAY DOSIMETRY
- Descriptors DEC
- ACCELERATORS; BEAMS; DISTRIBUTION; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATION DOSE DISTRIBUTIONS; RADIATIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS